The Impact of Methodological Differences on Static Lung Volume Measurement by Body Plethysmography
Maximum 25 words
Abstract Content
Body plethysmography (BPleth) measure thoracic gas volume (TGV) and allows estimation of lung volume subdivisions. The ERS/ATS 2023 guideline has recommended inspiratory capacity (IC) to derive total lung capacity (TLC) following panting manoeuvre against closed shutter. In practice, some patients struggle with this protocol and requires a spirometry separately. We would like to study the impact of unlinked spirometry on lung volume measurement. 
As part of quality improvement project (2023-2024), patients or bio-calibration subjects performed spirometry for slow vital capacity (SVC) and IC, prior to Bpleth. The largest values of ≥3 trials were obtained. Bpleth was then conducted using daily-calibrated MGC Diagnostics Platinum Elite Series Plethysmograph device, in accordance to ERS/ATS guidelines. Lung volume subdivisions obtained via ATS method were compared to those obtained via Bpleth TGV and unlinked spirometry.
Seventy-one individuals were identified. COPD (n=9), Asthma(n=6), and chronic cough (n=5) were the 3 main indications for Bpleth. Mean (standard deviation, SD) age is 61 (14) years; 46.5% were female and 85.9% were ethnic Chinese. Fifty-seven participants (80.3%) demonstrated good panting technique; though two could not perform IC manoeuvre in same setting. Five (7.1%) had very poor Bpleth technique, precluding analysis. Mean (SD) TLC was 4.96 (1.28) L, RV was 2.04 (0.66) L, IC was 2.07 (0.79), and SVC was 2.92 (0.98). These lung volume divisions measured by the 2 techniques were highly correlated (r = 0.90-0.98). Among individuals with good Bpleth technique, there was no significant differences in lung volume subdivisions between methods except for SVC (-0.08 [0.17] L, p-value <0.001). Nineteen of 57 (33%) had TLC difference of >5% between the technique. 
Combining Bpleth TGV and unlinked spirometry provides reasonably accurate estimates of lung volume subdivisions. However, clinically significant variability may occur. Results should therefore be interpreted with caution if same-setting IC cannot be performed during Bpleth. 
 
 
 
 
 